Thursday, April 27, 2017

Case #8 — A “De-Synching” Feeling

by David A. Burkland, MD, Mohammed Saeed, MD, FACC, and Yochai Birnbaum, MD, FACC


A 78-year-old man presented with worsening dyspnea and edema. He had undergone coronary artery bypass grafting in 2001. At the current presentation, he had ischemic cardiomyopathy with a left ventricular (LV) ejection fraction of 0.20 and was taking home inotropic therapy. Two months previously, he had begun cardiac resynchronization therapy with use of an implanted biventricular pacemaker.

Physical examination revealed elevated jugular venous pressure, bibasilar crackles, and pitting edema above both knees. An electrocardiogram (ECG) was obtained (Fig. 1).

Friday, February 17, 2017

Case #7 — Electrocardiogram Interpretation in a Man with Alcohol Withdrawal and Hypothermia

by Joanna Troulakis, MD, Roman Zeltser, MD, and Amgad N. Makaryus, MD

A 60-year-old man with paroxysmal atrial fibrillation, hypertension, seizure disorder, and alcohol abuse was unresponsive on presentation at the hospital. Laboratory tests revealed hypokalemia (2.5 mg/dL), hypomagnesemia (1.3 mg/dL), and no elevation in cardiac biomarkers. The patient’s admission electrocardiogram (ECG) showed an undetermined rhythm, with further interpretation limited by motion artifact. He was admitted with a diagnosis of alcohol withdrawal and hypothermia. During his hospital stay, he was monitored on telemetry for cardiac manifestations of electrolyte abnormalities. The covering physician was urgently called for suspicious telemetry events that prompted the completion of the following ECG (Fig. 1). 

Saturday, December 17, 2016

Case #6 — Proarrhythmic Antiarrhythmic

by Mark Pollet, MD, and Mohammad Saeed, MD

A 56-year-old man with a history of atypical atrial flutter related to atrial septal defect repair presented at the emergency department with palpitations and light-headedness. An electrocardiogram (ECG) was performed (Fig. 1). He had undergone an ablation procedure and therapy with metoprolol and flecainide. He had a history of cardiac arrest and had received a Teligen® model E110 dual-chamber implantable cardioverter-defibrillator (ICD) (Boston Scientific Corporation; Natick, Mass). The ICD was programmed in DDD mode, with a lower rate limit of 60 beats/min and an upper limit of 120 beats/min. Ventricular tachycardia therapy was set to begin at 210 beats/min, and ventricular fibrillation therapy at >230 beats/min.

Wednesday, October 12, 2016

Case #5 — To Pace or Not to Pace?

by Nadeen N. Faza, MD, Kainat Khalid, MD, Mohammad Saeed, MD, and Yochai Birnbaum, MD, FACC

A 75-year-old man with a medical history of sick sinus syndrome and left bundle branch block presented for evaluation of chest pain. The patient had a dual-chamber permanent pacemaker. Figure 1 shows his electrocardiogram (ECG).

Friday, August 5, 2016

Case #4 — An Unusual Presentation of QT Prolongation

by Mohammad Khalid Mojadidi, MD, Ninel Hovnanians, MD, Michael R. Kaufmann, MD, and James A. Hill, MD, FACC

A 55-year-old woman with a history of chronic bronchitis, Clostridium difficile colitis, and alcohol and tobacco abuse was admitted with altered mentation, hyponatremia, and necrotizing right-upper-lobe pneumonia. She was started on cefepime and vancomycin; metronidazole was added for colitis. A resting electrocardiogram (ECG) showed sinus tachycardia with a QS pattern in the precordial leads, normal intervals, and small T-wave inversions in leads V5 and V6. Four days later, significant changes in her baseline telemetry rhythm prompted a repeat 12-lead ECG; the patient’s pulse rate was 75 beats/min with a QT interval of 720 ms and QTc of 746 ms (Fig. 1). Her troponin T level was <0.03 ng/mL, and she had no new symptoms. Her medications at that time were aspirin, metoprolol, lisinopril, cefepime, atorvastatin, pantoprazole, metronidazole, oral vancomycin, and subcutaneous heparin. Her potassium level was 2.9 mEq/L, and her magnesium level was 1.6 mEq/L. An echocardiogram showed severe left ventricular dysfunction with wall motion that suggested stress-induced cardiomyopathy.

Friday, May 20, 2016

Case #3 - Heart Block in a Pacemaker: Does This Mean Trouble?

by Mark Pollet, MD, Yochai Birnbaum, MD, and Alireza Nazeri, MD

We present a finding in a 61-year-old woman whose St. Jude Medical dual-chamber permanent pacemaker had been implanted to treat symptomatic bradycardia. The pacemaker, programmed in DDD mode, had a lower rate limit of 60 beats/min and an upper limit of 120 beats/min, a paced atrioventricular (AV) delay period of 250 ms, and a sensed AV delay period of 225 ms.

Wednesday, April 6, 2016

Case #2 - Evaluation of Suspected Device Malfunction on ECG

by Luke Cunningham, MD, Henry D. Huang, MD, and Yochai Birnbaum, MD

A 62-year-old man with nonischemic cardiomyopathy, a history of Boston Scientific biventricular implantable cardioverter-defibrillator placement (in 2011), ventricular tachycardia after radiofrequency ablation (April and November 2014), paroxysmal atrial fibrillation, and severe mitral regurgitation presented with acute exacerbation of heart failure. A resting electrocardiogram (ECG) showed normal sequential atrioventricular (AV) pacing at a heart rate of 63 beats/min. Baseline device settings were DDD with a lower rate of 60 beats/min, an upper rate of 115 beats/min, a minimum sensed AV delay of 135 ms, and a minimum paced AV delay of 180 ms. The patient underwent mitral valve replacement. Three days later, the pacemaker rate was increased to 80 beats/min, and an ECG showed pacing concomitantly within the T wave in beats 5 and 13 of the rhythm strip (Fig. 1).