by Luke Cunningham, MD, and Yochai Birnbaum, MD
An 81-year-old woman with a
medical history of coronary artery disease, paroxysmal atrial fibrillation, and
sick sinus syndrome, and implantation in 2013 of an AltruaTM dual-chamber pacemaker (Boston
Scientific Corporation; Natick, Mass), presented after device interrogation
revealed elevated impedance of >2,500 Ω in her atrial lead. The device
settings were DDD mode, a lower rate limit of 60 beats/min, and a maximum atrioventricular
(AV) delay of 330 ms. She reported fatigue, dyspnea on exertion, and occasional
palpitations. She was admitted for new atrial lead implantation, and an
electrocardiogram was obtained (Fig. 1).
Fig. 1 |
The electrocardiogram shows
which of the following?
A) Ventricular
bigeminy
B) Normal ventricular-paced rhythm with undersensed premature atrial depolarizations
C) Normal ventricular pacing with underlying sinus rhythm
D) Normal ventricular pacing with retrograde atrial depolarizations
B) Normal ventricular-paced rhythm with undersensed premature atrial depolarizations
C) Normal ventricular pacing with underlying sinus rhythm
D) Normal ventricular pacing with retrograde atrial depolarizations
Answer
D) Normal ventricular pacing with retrograde atrial
depolarizations
Figure 1 shows ventricular depolarizations in a
bigeminal pattern in lead V1, all of which are preceded by a pacing
stimulus, which excludes selection A. In addition, the first beat (odd numbers)
in each pair is followed by an atrial depolarization (Fig. 2, arrows), which appears to be retrograde atrial
activation from the ventricular-paced beat.
Fig. 2 |
This observation is supported by the
predominantly negative deflection of the P waves and the proximity to the
preceding T wave. Therefore, underlying sinus rhythm is not seen, so selection
C is excluded. These atrial depolarizations appear to have been sensed by the
pacing device, because a ventricular-paced beat is seen after each atrial beat
at a consistent delay of 300 ms, and the programmed AV delay was 330 ms,
correlating with this finding. Selection B is therefore excluded.
The pattern of the tracing is of particular
note, given the grouped beats of ventricular pacing in a bigeminal pattern.
Absence of an atrial depolarization is noted after the second ventricular-paced
beat in each pair (even numbers). Therefore, second-degree 2:1 retrograde AV
block is present. Had retrograde conduction of ventricular-paced beats
continued, a pattern of pacemaker-induced tachycardia might have been seen.
“Endless-loop” tachycardia is a known sequela of DDD pacing in dual-chamber
pacemakers.1-3 Its characteristics are usually
a function of the set AV delay and ventriculoatrial (VA) blanking periods.3 Our
patient's programmed atrial blanking period after ventricular pacing was 120 ms
with an observed VA conduction of 320 ms, well outside this period. However,
pacemaker-induced tachycardia is not present, evidenced by the absence of
atrial depolarization after each second ventricular-paced beat.
In cardiac electrophysiologic studies of healthy
patients, VA conduction in ventricular pacing has had an average refractory
cycle length of 432 ms (range, 360–600 ms). Most patients (66%) had block
within the AV node, whereas the remaining patients had block within the
His-Purkinje system.4 Our patient was elderly and had
sick sinus syndrome, making underlying AV nodal disease highly likely. A longer
refractory period of the AV node could explain the intermittent absence of
retrograde P waves with ventricular pacing.
Ultimately, our patient underwent uncomplicated
implantation of a new right atrial lead, with normal device function
thereafter.
References
- Furman S, Fisher JD. Endless loop tachycardia in an AV universal (DDD) pacemaker. Pacing Clin Electrophysiol 1982;5(4):486–9.
- Tolentino AO, Javier RP, Byrd C, Samet P. Pacer-induced tachycardia associated with an atrial synchronous ventricular inhibited (ASVIP) pulse generator. Pacing Clin Electrophysiol 1982;5(2):251–9.
- Hayes DL, Furman S. Atrio-ventricular and ventriculo-atrial conduction times in patients undergoing pacemaker implant. Pacing Clin Electrophysiol 1983;6(1 Pt 1):38–46.
- Mahmud R, Denker S, Lehmann M, Gilbert C, Akhtar M. Functional characteristics of retrograde conduction in a pacing model of “endless loop tachycardia”. J Am Coll Cardiol 1984;3(6):1488–99.
No comments:
New comments are not allowed.