Alexander
Postalian, MD, Mohammad Saeed, MD, J. Alberto Lopez, MD, and Yochai Birnbaum,
MD
A 93-year-old
woman with no symptoms was admitted to the hospital for a pacemaker generator
exchange. She had a history of ischemic cardiomyopathy and was taking optimal
medical therapy. She had also experienced paroxysmal atrial arrhythmia, which
had been treated with atrioventricular node ablation and dual-chamber permanent
pacemaker implantation. Upon her arrival, a 12-lead electrocardiogram was
obtained (Fig. 1).
This ECG is most consistent with:
A) Sinus rhythm with a normally
functioning pacemaker set in DDD mode
B) Sinus rhythm with a pacemaker sensing
malfunction
C) Atrial tachycardia with a normally
functioning pacemaker set in DDD mode
D) Atrial tachycardia with a pacemaker
sensing malfunction
E) Atrial tachycardia with a normally
functioning pacemaker set in DDI mode
Answer
E)
Atrial tachycardia with a normally functioning pacemaker set in DDI mode
Fig. 2 |
Figure
2 shows atrial tachycardia and a normally functioning pacemaker set in DDI
mode.
Selections A and B are incorrect, because the P-wave morphology is not
consistent with sinus rhythm.
Selection
C is incorrect, because the paced R-R intervals are constant and the
atrioventricular intervals vary between QRS intervals. This signifies that
atrial depolarizations are not triggering ventricular pacing, which would occur
in DDD mode.
Selection
D is incorrect, because the pacemaker is appropriately sensing atrial activity.
During
most cardiac cycles, only the ventricle is being paced, regardless of atrial
activity. The reason is atrial channel inhibition by native atrial impulses
(black arrows).
However,
there are 2 atrial-paced beats (arrowheads). These atrial pacing events occur
because the native atrial beats preceding them (white arrows) came during the postventricular
atrial refractory period (PVARP), and no other P wave was sensed before the
ventriculoatrial or atrial escape interval timed out. The pacemaker does not
trigger when atrial activity takes place during the PVARP. The black arrows
point to the rest of the native atrial impulses.
Our
conclusion was confirmed on the pacemaker’s intracardiac electrogram. With the
PVARP set at 400 ms and the rate at 70 beats/min, atrial sensing and infrequent
atrial pacing were noted. Increasing the rate to 85 beats/min caused the PVARP
to cover a larger proportion of the R-R interval. Thus, native atrial impulses fell
within the PVARP more frequently, failing to inhibit atrial output, so the
frequency of intermittent atrial pacing increased.
The DDI mode can
be used to avoid tracking atrial arrhythmias. In this mode, the pacemaker is
inhibited by atrial beats that would otherwise trigger a ventricular event in
DDD mode. Most modern pacemakers automatically switch from DDD to DDI mode if
atrial tachyarrhythmias are detected. This feature is called mode switch.
No comments:
New comments are not allowed.