Friday, February 15, 2019

Case #17 — Tachycardia in the Presence of Ventricular Pacing


by Toug Tanavin, MD, Mark Pollet, MD, and Yochai Birnbaum, MD, FACC


A 61-year-old man with coronary artery disease presented with volume overload caused by medication noncompliance. His medical history included percutaneous coronary intervention; ischemic cardiomyopathy (left ventricular [LV] ejection fraction, <0.15); and placement of a biventricular implantable cardioverter-defibrillator (ICD), model and programming unknown. Chest radiographs showed properly positioned ventricular and right atrial leads. Figure 1 shows the patient’s electrocardiogram (ECG) on presentation.


Fig. 1

On the basis of the ECG, what is the patient’s diagnosis?
A) Supraventricular tachycardia with biventricular pacing response
B) Pacemaker-mediated tachycardia
C) Sinus tachycardia with normal atrioventricular delay programming
D) Antitachycardia pacing



Answer

A) Supraventricular tachycardia with biventricular pacing response

The ECG shows a relatively narrow QRS complex at a ventricular rate of 100 beats/min. Notably, an rSR' configuration in leads V1 and V2 (Fig. 2A, arrows) represents a retrograde P wave that suggests supraventricular tachycardia—most likely atrioventricular nodal reentry tachycardia, not sinus tachycardia with normal atrioventricular delay programming.1 The ventricular pacemaker stimuli after each intrinsic QRS complex are probably from responsive biventricular ICD pacing. After the patient’s tachycardia resolved, we observed sinus rhythm, normal right bundle branch configuration, and normal biventricular pacing (Fig. 2B).

Fig. 2A

Fig. 2B

When the ICD senses intrinsic activity in one ventricle, it immediately paces the other ventricle, to better synchronize LV contraction. However, the initially narrower QRS configuration suggests that the LV pacing stimuli were not capturing the already depolarized ventricular myocardium. Responsive biventricular ICD pacing is expected to produce fusion complex morphology, as in cardiac resynchronization therapy, and indeed the wider QRS complex after conversion to sinus rhythm indicates fusion from His–Purkinje system depolarization with coinciding pacing stimuli.

In Figure 1, the response is not antitachycardia pacing, because the pacing stimuli occur before the QRS complexes; moreover, there was no burst of pacing stimuli faster than the spontaneous rate.1 Nor is the response pacemaker-mediated tachycardia, because the pacing stimuli occur after the QRS complexes begin.2,3 The P waves at the end of the QRS complexes might be retrograde conduction from the pacemaker stimulus; however, the very long subsequent PR intervals suggest another mechanism.

References
  1. Issa ZF, Miller JM, Zipes DP. Clinical arrhythmology and electrophysiology: a companion to Braunwald’s heart disease. 2nd ed. Philadelphia: Elsevier Saunders; 2012. p. 381-410. 
  2. Cha YM, Mulpuru SK. Cardiac resynchronization therapy programming and troubleshooting. In: Ellenbogen KA, Wilkoff BL, Kay GN, Lau CP, Auricchio A, editors. Clinical cardiac pacing, defibrillation, and resynchronization therapy. 5th ed. Philadelphia: Elsevier; 2017. p. 1090-132. 
  3. Burri H. Pacemaker programming and troubleshooting. In: Ellenbogen KA, Wilkoff BL, Kay GN, Lau CP, Auricchio A, editors. Clinical cardiac pacing, defibrillation, and resynchronization therapy. 5th ed. Philadelphia: Elsevier; 2017. p. 1031-63.
https://doi.org/10.14503/THIJ-18-6810

No comments: